Effect of different kinematics and operational temperature on cyclic fatigue resistance of rotary NiTi systems

Abstract

Background

The aim of this study was to compare the effect of different kinematics and operational temperature on cyclic fatigue resistance of Protaper Next, WaveOne Gold, and Twisted File Adaptive files in a static model.

Methodology

An artificial canal, made of stainless steel with an inner diameter of 1.5 mm, a 60° angle of curvature, and a curvature radius of 2 mm, was used for the cyclic fatigue test immersed in a water bath for temperature adjustment (37 °C and 20 °C). A total of 120 files were divided into three groups according to the type of motion continuous rotation motion (CRM), adaptive motion (AM), and reciprocation motion (RM). Each group was further subdivided into two subgroups according to the operational temperature (37 °C and 20 °C), All instruments were operated until fracture occurred, and the time to fracture was recorded in seconds. Mean number of cycles to failure and standard deviations were calculated for each group, and data were analyzed statistically using two-way ANOVA followed by Tukey’s post hoc test (P ≤ 0.05).

Results

The continuous rotation motion exhibited significantly lower cyclic fatigue resistance than reciprocating or adaptive motion (p ≤ 0.001) at 37 °C and 20 °C.

Conclusions

Cyclic fatigue resistance of NiTi files was affected by the motion used during root canal preparation. Reciprocation motion and adaptive motion resulted in extended fatigue. The operational temperature affects the cyclic fatigue resistance of the twisted file.

Introduction

The introduction of rotary nickel-titanium (NiTi) instruments in endodontic practice has revolutionized the concept of root canal instrumentation in the last two decades. This is because of the alloy’s superelasticity associated with the reversible phase transformation between austensite and martensite phases (Shen et al. 2013; Pereira et al. 2012; Brantley et al. 2009).

The clinical behavior of these two phases is remarkably different and temperature dependent (Otsuka and Ren 2005). Originally, the development of NiTi systems has been focused on changes in file design together with the simplification of the instrumentation sequences without changing the basic properties of the alloy (Shen et al. 2013; Peters and Paque 2010). More recently, new heat treatment methods together with novel manufacturing procedures have been introduced to improve the cyclic fatigue resistance of the rotary files (Plotino et al. 2017; Gambarini et al. 2008; Rodrigues et al. 2011).

Another approach adopted by the manufacturers to improve the cyclic fatigue resistance of the instruments is changing the rotation kinematics during root canal preparation (Çapar and Arslan 2016). Reciprocation motion was introduced to reduce stress values attained by the instrument during rotation through travelling to a shorter angular distance than rotation motion do (Silva et al. 2016; Plotino et al. 2015). Meanwhile, the adaptive motions is composed of both continuous and reciprocation motion. When the amount of stress on the file is kept to minimal, the file continues in constant rotation until substantial load is applied; hence, reciprocation motion is executed (Marks Duarte et al. 2018).

Fracture of endodontic rotary instruments occurs due to combination between torsional and flexural fatigue. Torsional fracture occurs due to locking of the tip or any part of the instrument in the canal, while the shaft continues to rotate. Flexural fatigue occurs when the instrument rotates inside a curved root canal and is subjected to an excessive number of tension-compression strain cycles in the region of maximum root canal curvature (Pedullà et al. 2018). The environmental temperature dramatically affects fatigue resistance of nickel-titanium rotary files (Plotino et al. 2017; Grande et al. 2017).

The aim of the present study was to compare the effect of different kinematics on cyclic fatigue resistance of Protaper Next (PTN), WaveOne Gold (WOG), and Twisted File Adaptive (TFA) files in room temperature.

Materials and methods

Based on data from previous studies (Plotino et al. 2017; De Vasconcelos et al. 2016), a power calculation was performed using the G*Power 3.1 software (Faul et al. 2007) (Heinrich Heine University, Dusseldorf, Germany). The calculation indicated that the sample size for each group should be a minimum of 20 files. A total of 120 files were tested for cyclic fatigue failure, twenty files from each system; TFA; SM2 (SybronEndo, Orange, CA, USA), PTN; X2 (Dentsply Maillefer, Ballaigues, Switzerland); and WOG; primary file (Dentsply Maillefer, Ballaigues, Switzerland). The instruments were inspected for any manufacturing defects using a stereomicroscope (Olympus BX43; Olympus Co, Tokyo, Japan).

A specially designed custom-made cyclic fatigue testing device was used in this study (Fig. 1). This device consisted of curved stainless steel canal 19 mm in length with 60° angle of curvature and 2 mm radius, which were stuck on an acrylic surface where the headpiece was also fixed. The radius and angle of curvature were determined according to Pruett et al. (Pruett et al. 1997). To reduce friction between the instrument and the metal canal walls, a synthetic oil was used as a lubricant. The artificial canals were covered with glass to prevent the instruments from slipping out.

Fig. 1
figure 1

A custom-made cyclic fatigue testing device

The method described by De Vasconcelos et al. (De Vasconcelos et al. 2016) was used to adjust the operating temperature. The device was placed in a glass water container, with rotating the instruments until fracture in air. The glass container was filled with 200 mL water and immersed in ice until the water temperature was stabilized at 20 ± 1 °C or placed on a hot plate until the water temperature was stabilized at 37 ± 1  °C, and the temperature was continuously monitored using an infrared thermometer (Etekcity Corporation, CA, USA).

The files were divided into three groups (n = 20), and the following procedures were performed:

Group (RM): reciprocation motion

The WOG primary files 25/.07 (n = 20) were inserted into the artificial canal and operated using a torque-controlled endodontic motor (x-smart plus, Dentsply, Maillefer) following the instructions given by the manufacturer at 350 rpm, reciprocating motion until fracture.

Group (AM): adaptive motion

The TFA files SM2 25/.06 (n = 20) were inserted into the artificial canal and operated with the “TF Adaptive” program, using electric motor (SybronEndo, Glendora, CA, USA) following the instructions given by the manufacturer at 400 rpm until fracture.

Group (CRM): continuous rotation motion

The PTN X2 files (n = 20) were inserted into the artificial canal and operated using a torque-controlled endodontic motor (x-smart plus, Dentsply, Maillefer) following the instructions given by the manufacturer at 300 rpm, continuous rotation motion until fracture.

The time to fracture was recorded in seconds using a stopwatch and stopped as soon as a fracture was detected visually, and a video recording was performed to cross-check the time of file separation.

The number of cycles to failure (NCF) was calculated by multiplying the rpm (in seconds) with the time taken to fracture (in seconds). NCF data were analyzed by using one-way ANOVA followed by Tukey’s post hoc test. Statistical analysis was performed using the software SPSS (Statistical Packages for the Social Sciences 20.0; IBM, Armonk, NY).

Results

The mean NCF and standard deviations for all groups are shown in Table 1.

Table 1 The mean and standard deviation (SD) of the number of cycles to failure (NCF) for different groups

There was a statistically significant difference between (AM), (RM), and (CRM) (p ≤ 0.001) at 37 °C and 20 °C.

At 20 °C, the highest fatigue resistance was recorded in group (RM) Wave One gold followed by group (AM), TF adaptive without significant difference, whereas the lowest fatigue resistance was recorded in group (CRM) Protaper Next which was significantly lower.

At 37 °C, the highest fatigue resistance was recorded in group (RM) Wave One Gold followed by group (AM), TF adaptive with significant decrease, whereas the lowest fatigue resistance was recorded in group (CRM) Protaper Next which was significantly lower.

Regarding the effect of temperature, the RM group (Wave One Gold) and RCM group (Protaper Next) showed no significant decrease in NCF with raising the temperature to 37 °C. However, the AM group (Twisted file) recorded significant decrease in NCF at 37 °C.

Discussion

Several factors affect NiTi file fracture during endodontic treatment of curved root canals such as kinematics, speed used, structural characteristics, method of manufacturing, geometric designs, radius, and angle of the curvature of the canal (Castelló-Escrivá et al. 2012; Kim et al. 2008).

The aim of the present study was to compare the effect of different kinematics on cyclic fatigue resistance of Protaper Next, WaveOne Gold, and TF Adaptive files in a static model at different operational temperatures.

According to Yao et al. (Yao et al. 2006), to decrease the influence of other variables, standardized artificial canals in custom made device should be used. Therefore, in the present study, a standardized artificial canal was used for the cyclic fatigue test, and this was in full agreement with De-Deus et al. (De-Deus et al. 2010).

A static model was used in this study rather than a dynamic model. Although the dynamic model could mimic the clinical brushing or pecking motion, it has limitations because the instruments are not constrained in a precise trajectory. Moreover, the speed and amplitude of the axial movements are subjective because this up and down motion is manually controlled (Pedullà et al. 2018; Higuera et al. 2015).

The results of this study showed that the highest cyclic fatigue resistance was associated with the reciprocation motion followed by adaptive motion then continuous rotation. This might indicate that the motion plays a major role in cyclic fatigue resistance. These findings might be attributed to the release of the reaction stresses built up in the material by reversing the rotational direction (Gavini et al. 2012; You et al. 2010). Plotino et al. (Plotino et al. 2014) concluded that reciprocating movement can improve cyclic fatigue resistance of NiTi instruments by reducing torsional loads and torsional failure.

On the other hand, the adaptive motions are composed of continuous and reciprocation motion. When the amount of stress on the file is minimal, the file uses continuous rotation till load is applied, and it then uses reciprocation motion (Higuera et al. 2015); thus, TFA has been shown to be elastic and resistant to cyclic fatigue than continuous rotation motion did (De-Deus et al. 2010; Bhagabati et al. 2012).

A number of studies (Gambarini et al. 2012; Karataş et al. 2016; Keskin et al. 2017; Gündoğar and Özyürek 2017) reported that, regardless of the file type, reciprocation motion increased the cyclic fatigue resistance of files more than continuous rotation did. Karatas et al. (Karataş et al. 2016) concluded that fatigue life is associated with the number of times that a crack in an instrument closes and opens. During reciprocation motions, the instrument needs to complete one full rotation (360) in more time when compared with continuous rotation. Therefore, the number of times that a crack closes and opens in reciprocation motions is expected to be less when compared with continuous rotation or adaptive motions over the same time. This may explain the extended fatigue life of reciprocation motion. This was in full agreement with the results of the present study as the PTN showed the least cyclic resistance among the tested groups.

WOG file has a parallelogram cross-sectional design with only two cutting edges, and this design in addition to high austenitic finish value and 2-stage transformation behavior decreases the torsional stress on the file and extends its cyclic fatigue resistance as described by Özyürek (Gündoğar and Özyürek 2017; Özyürek 2016). Although TFA file has a triangular cross-section design with three cutting edges that thought to increase the torsional stress on the file decreasing its cyclic fatigue resistance, yet, the unique heat treatment (R-phase) coupled with electropolishing (Castelló-Escrivá et al. 2012) might maintain the natural grain structure resulted in decreased crack initiation and propagation process. Hence, the file has comparable cyclic fatigue resistance coupled with the novel adaptive motion operational mode.

Although PTN file has the same parallelogram cross-section design as WOG, it showed significant difference in cyclic fatigue resistance than WOG and TFA files. These results emphasized that the motion plays a major role in cyclic fatigue resistance than the design do.

Another important issue to be emphasized is the operational temperature. It was found that the temperature inside the root canal is buffered within few minutes to 31–35 °C as stated by De Hemptinne et al. (De Hemptinne et al. 2015). So, our study was performed comparing the two environmental temperatures. It was found that raising the temperature from room temperature to body temperature considerably reduced the cyclic fatigue resistance for both PTN and TFA rotary systems. These findings are in agreement with recent previous studies (Plotino et al. 2017; Grande et al. 2017; De Vasconcelos et al. 2016; Yılmaz et al. 2018).

The considerable decrease in fatigue resistance might be due to the interrelated transformation temperatures, austenitic finish temperature, in which raising the temperature to around 37 °C suggests that the instrument is typically in an austenitic state. This is evident for TFA in which the austenitic finish temperature was identified to be about 20 °C (Braga et al. 2013), whereas in Wave One Gold and the M-Wire Protaper Next, the transformation temperature obviously above body temperature is about 50 °C (Ye and Gao 2012; Hieawy et al. 2015); accordingly, the files were not affected by the increase of temperature during cyclic fatigue testing in the present study which was in full agreement with Plotino et al. (Plotino et al. 2017).

Conclusion

Under the conditions of this study, it can be concluded that:

  1. 1.

    The operational temperature at which the instruments are tested has a considerable effect on cyclic fatigue testing and varies according to the type of alloy.

  2. 2.

    Gold heat treated alloys and M-wire alloys are not readily affected by the intracanal temperature because their austenitic temperature is obviously above body temperature, whereas Twisted file is influenced by intracanal temperature because its austenitic finish is below body temperature.

  3. 3.

    The mode of rotation greatly affects the cyclic fatigue, whereas both the reciprocation and adaptive motions exhibited higher cyclic fatigue resistance than continuous rotation.

Availability of data and materials

• The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

• All data generated or analyzed during this study are included in this published article.

Abbreviations

NiTi:

Nickel titanium

PTN:

Protaper Next

WOG:

WaveOne Gold

TFA:

Twisted File Adaptive

SPSS:

Statistical Packages for social sciences

ANOVA:

Analysis of variance

CRM:

Continuous rotation motion

AM:

Adaptive motion

RM:

Reciprocating motion

NCF:

Number of cycles to failure

RPM:

Revolutions per minute

References

  • Bhagabati N, Yadav S, Talwar S (2012) An in vitro cyclic fatigue analysis of different endodontic nickel-titanium rotary instruments. J Endod. 38(4):515–518

    Article  Google Scholar 

  • Braga LC, Magalhães RRS, Nakagawa RKL, Puente CG, Buono VTL, Bahia MGA (2013) Physical and mechanical properties of twisted or ground nickel-titanium instruments. Int Endod J. 46(5):458–465

    Article  CAS  Google Scholar 

  • Brantley WA, Liu J, Clark WAT, Kovarik L, Buie C, Iijima M, et al. Characterization of new nickel-titanium wire for rotary endodontic instruments. In: Ceramic Transactions. 2009. p. 49–57.

  • Çapar ID, Arslan H (2016) A review of instrumentation kinematics of engine-driven nickel-titanium instruments. International Endodontic Journal 49:119–135

    Article  Google Scholar 

  • Castelló-Escrivá R, Alegre-Domingo T, Faus-Matoses V, Román-Richon S, Faus-Llácer VJ (2012) In vitro comparison of cyclic fatigue resistance of ProTaper, WaveOne, and twisted files. J Endod. 38(11):1521–1524

    Article  Google Scholar 

  • De Hemptinne F, Slaus G, Vandendael M, Jacquet W, De Moor RJ, Bottenberg P (2015) In vivo intracanal temperature evolution during endodontic treatment after the injection of room temperature or preheated sodium hypochlorite. J Endod. 41(7):1112–1115

    Article  Google Scholar 

  • De Vasconcelos RA, Murphy S, Carvalho CAT, Govindjee RG, Govindjee S, Peters OA (2016) Evidence for reduced fatigue resistance of contemporary rotary instruments exposed to body temperature. J Endod. 42(5):782–787

    Article  Google Scholar 

  • De-Deus G, Moreira EJL, Lopes HP, Elias CN (2010) Extended cyclic fatigue life of F2 ProTaper instruments used in reciprocating movement. Int Endod J. 43(12):1063–1068

    Article  CAS  Google Scholar 

  • Faul F, Erdfelder E, Lang AG, Buchner A. G*Power 3: A flexible statistical power analysis program for the social, behavioral, and biomedical sciences. In: Behavior Research Methods. 2007. p. 175–91.

  • Gambarini G, Grande NM, Plotino G, Somma F, Garala M, De Luca M et al (2008) Fatigue Resistance of engine-driven rotary nickel-titanium instruments produced by new manufacturing methods. J Endod. 34(8):1003–1005

    Article  Google Scholar 

  • Gambarini G, Gergi R, Naaman A, Osta N, Al SD (2012) Cyclic fatigue analysis of twisted file rotary NiTi instruments used in reciprocating motion. Int Endod J. 45(9):802–806

    Article  CAS  Google Scholar 

  • Gavini G, Caldeira CL, Akisue E, Candeiro GTDM, Kawakami DAS (2012) Resistance to flexural fatigue of reciproc R25 files under continuous rotation and reciprocating movement. J Endod. 38(5):684–687

    Article  Google Scholar 

  • Grande NM, Plotino G, Silla E, Pedullà E, DeDeus G, Gambarini G et al (2017) Environmental temperature drastically affects flexural fatigue resistance of nickel-titanium rotary files. J Endod. 43(7):1157–1160

    Article  Google Scholar 

  • Gündoğar M, Özyürek T (2017) Cyclic fatigue resistance of OneShape, HyFlex EDM, WaveOne Gold, and Reciproc Blue Nickel-titanium instruments. J Endod. 43(7):1192–1196

    Article  Google Scholar 

  • Hieawy A, Haapasalo M, Zhou H, Wang ZJ, Shen Y (2015) Phase transformation behavior and resistance to bending and cyclic fatigue of ProTaper Gold and ProTaper Universal instruments. J Endod. 41(7):1134–1138

    Article  Google Scholar 

  • Higuera O, Plotino G, Tocci L, Carrillo G, Gambarini G, Jaramillo DE (2015) Cyclic fatigue resistance of 3 different nickel-titanium reciprocating instruments in artificial canals. J Endod. 41(6):913–915

    Article  Google Scholar 

  • Karataş E, Arslan H, Büker M, Seçkin F, Çapar ID (2016) Effect of movement kinematics on the cyclic fatigue resistance of nickel-titanium instruments. Int Endod J. 49(4):361–364

    Article  Google Scholar 

  • Keskin C, Inan U, Demiral M, Keleş A (2017) Cyclic Fatigue Resistance of Reciproc Blue, Reciproc, and WaveOne Gold Reciprocating Instruments. J Endod. 43(8):1360–1363

    Article  Google Scholar 

  • Kim HC, Cheung GSP, Lee CJ, Kim BM, Park JK, Kang S (2008) Il. Comparison of forces generated during root canal shaping and residual stresses of three nickel-titanium rotary files by using a three-dimensional finite-element analysis. J Endod. 34(6):743–747

    Article  Google Scholar 

  • Marks Duarte P, Barcellos da Silva P, Alcalde MP, Vivan RR, da Rosa RA, Duarte MAH et al (2018) Canal transportation, centering ability, and cyclic fatigue promoted by Twisted File Adaptive and Navigator EVO instruments at different motions. J Endod. 44(9):1425–1429

    Article  Google Scholar 

  • Otsuka K, Ren X (2005) Physical metallurgy of Ti-Ni-based shape memory alloys. Prog Mater Sci. 50(5):511–678

    Article  CAS  Google Scholar 

  • Özyürek T (2016) Cyclic fatigue resistance of Reciproc, WaveOne, and WaveOne Gold Nickel-Titanium instruments. J Endod. 42:1536–1539

    Article  Google Scholar 

  • Pedullà E, Corsentino G, Ambu E, Rovai F, Campedelli F, Rapisarda S et al (2018) Influence of continuous rotation or reciprocation of Optimum Torque Reverse motion on cyclic fatigue resistance of nickel-titanium rotary instruments. Int Endod J. 51(5):522–528

    Article  Google Scholar 

  • Pereira ESJ, Peixoto IFC, Viana ACD, Oliveira II, Gonzalez BM, Buono VTL et al (2012) Physical and mechanical properties of a thermomechanically treated NiTi wire used in the manufacture of rotary endodontic instruments. Int Endod J. 45(5):469–474

    Article  CAS  Google Scholar 

  • Peters O, Paque F (2010) Current developments in rotary root canal instrument technology and clinical use: a review. Quintessence Int. 41(6):479–488

    PubMed  Google Scholar 

  • Plotino G, Giansiracusa Rubini A, Grande NM, Testarelli L, Gambarini G (2014) Cutting efficiency of reciproc and waveone reciprocating instruments. J Endod. 40(8):1228–1230

    Article  Google Scholar 

  • Plotino G, Ahmed HMA, Grande NM, Cohen S, Bukiet F (2015) Current assessment of reciprocation in endodontic preparation: a comprehensive review - Part II: Properties and Effectiveness. Journal of Endodontics 41:1939–1950

    Article  Google Scholar 

  • Plotino G, Grande NM, Mercadé Bellido M, Testarelli L, Gambarini G (2017) Influence of temperature on cyclic fatigue resistance of ProTaper Gold and ProTaper Universal Rotary Files. J Endod. 43(2):200–202

    Article  Google Scholar 

  • Pruett JP, Clement DJ, Carnes DL (1997) Cyclic fatigue testing of nickel-titanium endodontic instruments. J Endod. 23(2):77–85

    Article  CAS  Google Scholar 

  • Rodrigues RCV, Lopes HP, Elias CN, Amaral G, Vieira VTL, De Martin AS (2011) Influence of different manufacturing methods on the cyclic fatigue of rotary nickel-titanium endodontic instruments. J Endod. 37(11):1553–1557

    Article  Google Scholar 

  • Shen Y, Zhou HM, Zheng YF, Peng B, Haapasalo M (2013) Current challenges and concepts of the thermomechanical treatment of nickel-titanium instruments. J Endod. 39(2):163–172

    Article  Google Scholar 

  • Silva EJNL, Villarino LS, Vieira VTL, Accorsi-Mendonça T, Antunes HDS, De-Deus G et al (2016) Bending resistance and cyclic fatigue life of Reciproc, Unicone, and WaveOne Reciprocating instruments. J Endod. 42(12):1789–1793

    Article  Google Scholar 

  • Yao JH, Schwartz SA, Beeson TJ (2006) Cyclic fatigue of three types of rotary nickel-titanium files in a dynamic model. J Endod. 32(1):55–57

    Article  Google Scholar 

  • Ye J, Gao Y (2012) Metallurgical characterization of M-Wire nickel-titanium shape memory alloy used for endodontic rotary instruments during low-cycle fatigue. J Endod. 38(1):105–107

    Article  Google Scholar 

  • Yılmaz K, Uslu G, Gündoğar M, Özyürek T, Grande NM, Plotino G (2018) Cyclic fatigue resistances of several nickel-titanium glide path rotary and reciprocating instruments at body temperature. Int Endod J. 51(8):924–930

    Article  Google Scholar 

  • You SY, Bae KS, Baek SH, Kum KY, Shon WJ, Lee W (2010) Lifespan of one nickel-titanium rotary file with reciprocating motion in curved root canals. J Endod. 36(12):1991–1994

    Article  Google Scholar 

Download references

Acknowledgements

Not applicable

Funding

None

Author information

Authors and Affiliations

Authors

Contributions

Amira Galal Ismail prepared specimens and analyzed and interpreted data. Manar Galal wrote the manuscript. Mohamed Mokhtar Nagy analyzed the data and substantively revised the work. The authors read and approved the final manuscript.

Authors’ information

Dr. Amira Galal is a graduate and postgraduate (Msc, phD) of Cairo University, School of Oral and Dental Medicine. She is currently a researcher of Endodontics, Restorative and Dental Material department at the National Research Centre. Dr. Amira is also an FDI speaker and has lectured extensively nationally and internationally. Moreover, she has published articles in local and international Journals.

Corresponding author

Correspondence to Amira Galal Ismail.

Ethics declarations

Ethics approval and consent to participate

Not applicable

Consent for publication

Not applicable

Competing interests

No competing interest in this section.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Ismail, A.G., Galal, M. & Nagy, M.M. Effect of different kinematics and operational temperature on cyclic fatigue resistance of rotary NiTi systems. Bull Natl Res Cent 44, 116 (2020). https://doi.org/10.1186/s42269-020-00370-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s42269-020-00370-4

Keywords